Phaco system / Phaco unit MD-480A

All parts of the system are fully controlled and monitored by Microprocessor Technology.
Smoothly and efficiently controlling the emulsifier process with digital frequency tracing technology of handpiece
Easy operation and user-fiendly interface with large touch screen TFT and functional menu
Up to 10 sets of parameters preset meets diferent requirements and simplifies the surgery operation-
Auto test checks tubing and handpiece problems (e.g.: block or leakage)
The peristaltic pump gives complete control of aspiration flow rate as well as freedom from anterior chamber shallowing. This way can give stronger sution when working in 1/A
Built-in vitrectomy and cautery.
Intelligent voice reminder

Parameter Specification
Ultrasonic Emulsification (U/S)
Frequency40 kHz
Functional modesContinuance, Pulse, Linear, Burst
Continuance (power range)0–100%
Pulse (frequency range)1–99 Hz
Linear (control)Power linearly controlled by foot switch, 0–100%
Burst stopping time1.5–0 s
Working time5–100 ms
Maximum vacuum500 mmHg
Maximum flux rate40 ml/min
Irrigation / Aspiration (I/A)
Flux rate0–40 ml/min
Reflux rate20 ml/min
Max vacuum500 mmHg
Vitrectomy (VIT)
Cutting power sourceAir pump
Cutting frequency20–600 / min
ModeSingle cut
Maximum vacuum500 mmHg
Maximum flux rate40 ml/min
Cautery
Maximum power10 W ± 20% (200 Ω)
TypeBipolar
Power preset range0–100%
Scroll to Top